Oxygen analyzer for wave soldering and reflow soldering can be widely applied in the online oxygen analysis in protection gas from Wave-soldering machine, reflow-soldering machine, electron,welding furnace for semiconductor industry and sintering furnace for inactive gas, etc.
Oxygen analyzer has friendly communication menu which makes the operation visual and simple. It could keep working in the air for a long time without affecting the sensor life. The response speed is rapid and the lowest measurable unit is 0.1ppm. Power being supplied at a wide range makes it more suitable for different occasions.
1.The imported zirconium dioxide sensor is with high measurement accuracy, a rapid response speed and long calibration cycle for features;
2.Only one random point needs to be calibrated among the whole measured range, making the calibration operation simple and convenient;
3.Function of built-in compensation system for temperature could reduce the influence from the changes of environment temperature on the measured accuracy;
4.No reference gas is needed, and the analyzer wouldn’t be affected by the oxygen concentration of working around atmosphere;
5.It could do one-way or bidirectional communication with the upper computer;
6.The imported sampling pump is built in and the working mode can be set through the software.
Test principle | zirconium dioxide |
Display mode | 128×64 matrix dot LCD |
Test range | 0 ~ 10/100/1000ppm / 1.00% / 25.00% O2 |
Resolution | 1ppm |
Accuracy | 0~1000ppm /1.00% / 10.00% / 25.00% ≤±1% FS 0~100ppm ≤±2% FS 0~10ppm ≤±5%FS |
Repeatability | ≤±1% |
Response time | T90≤30S |
Analog output | 4~20mA.DC(non-isolated output, loaded resistance lower than 500 Ohm) 0~10V.DC(non-isolated output, loaded resistance higher than 10k Ohm) 1 way programmable alarm output (Dry contact type, maximum load 220VAC/2A) |
Communication port | RS232 (default) or RS485 |
Power supply | AC85 ~ 264V 50/60Hz |
Temperature of the sample gas | -10 ~ +50℃ |
Sample gas flow | 400 ~ 600ml/min |
Sampling mode | extraction way, ejection way, extraction and ejection way |
Sample gas pressure | micro positive pressure, micro negative pressure or normal pressure |
Background gas | N2 and inactive gaseous mixture |
Appearance Size | 133mm×268mm×300mm(H×W×D) |
Install Hole Cut | 135mm×228mm(H×W) |
Serving life | >4 years(under normal use) |
Gas circuit port | NPT 1/8 internal thread |
Install mode | frame mount |
Disclaimer: The product introduction content (including product images, product descriptions, technical parameters, etc.) of this site is for reference only. As the update may not be timely, there will be some differences between the content and the actual situation. Please contact our sales staff to confirm. The information provided on this site does not constitute any offer or promise, and the company will improve and modify any information on the website from time to time without prior notice.